SPECIFICATIONS
Technical:
Insulation: Class2, Double insulation.
Overvoltage category: CATIV 600V,CATIII 1000V
NOTE: These meters meet CAT III and CAT IV IEC 61010
standards. The IEC 61010 safety standard defines four
overvoltage categories (CAT I to IV) based on the magnitude
of danger from transient impulses. CAT III meters are
designed to protect against transients in fixed-equipment
installations at the distribution level; CAT IV meters are
designed to protect against transients from the primary
supply level (overhead or underground utility service).
Maximun voltage between any terminal and earth
ground: 1000V DC/AC RMS
Surge Protection: 8kV peak IEC 61010
Display: 4000 counts LCD display, 21mm high
Polarity: Automatic, (-) negative polarity indication.
Over-range: “OL” mark indication.
Low battery indication: A battery “ “symbol is
displayed when the battery voltage
drops below the operating level.